This article applies to:
E-Prime Extensions for Net Station 3.0
E-Prime Extensions for Net Station 2.0
Symptoms
A runtime error occurs when specifying a Subject number greater than 32767:
The following runtime error occurred:
Overflow
Error Number: 6
Detail
Once the runtime error is dismissed, the Output window states the following:
ERROR: The error occurred in the method named "NetStation_WriteSessEvent(c" at line number 92 of a Certified Package File
The following runtime error occurred:
Workaround
Workaround 1:
If specifying Subject numbers greater than 32767 is required then create an additional StartupInfo parameter (e.g., MyCustomSubject#), set Data Type to String rather than Numeric or Choice, and set the Max Length to allow for enough characters.
NOTE: When running the experiment, make sure to set E-Prime's default Subject number parameter is greater than 1 to ensure data collection (Subject number 0 = no data collection).
Workaround 2:
For E-Prime Extensions for Net Station enabled experiments, do not exceed Subject number 32767.
See Also:
RELEASE INFO: E-Prime Extensions for Net Station 3.0 Known Issues [27227]